The DMC 3000 Electronic Dosimeter is a personal radiation dosimeter for health physics and radiation protection personnel that detects X-ray and Gamma radiation with audible, visual, and tactile alarms.
The DMC 3000 Electronic Dosimeter is a personal radiation dosimeter designed for health physics and radiation protection personnel. It provides detection and measurement of X-ray and Gamma radiation with superior energy response (Hp(10) and Hp(0.07)). Features include a high-contrast backlit LCD display, multiple alarm methods (audible, visual, and tactile), simple 2-button navigation, and operates for over 3,000 hours on a single AAA alkaline battery. The device exceeds applicable IEC and ANSI standards and offers excellent electromagnetic interference immunity. Additional modules are available for Beta, Neutron, and Telemetry capabilities. Compatible with Vital Supervision software and integrates with contamination monitoring systems.
| Performance | |
| Temperature range | -10 °C to +50 °C (+14 °F to +122 °F) |
| Alarming speaker | level of 85 dB (A) typical (> 90 dB (C) peak) at 30 cm (11.8 in), frequency < 4800 Hz |
| Dose increments resolution | 1 μSv (0.1 mrem) resolution |
| Hp(10) γ Effective range of dose | 0.01 μSv to 100 Sv (0.001 mrem to 10000 rem) |
| Hp(10) γ Display resolution | 0.1 μSv to 10 Sv (0.01 mrem to 1000 rem) up to four decimal places |
| Hp(0.07) γ Effective range of dose | 0.01 μSv to 100 Sv (0.001 mrem to 10000 rem) |
| Hp(0.07) γ Display resolution | 0.1 μSv to 10 Sv (0.01 mrem to 1000 rem) up to four decimal places |
| Hp(10) γ Effective range of dose rate | 0.05 μSv/h to 20 Sv/h (0.005 mrem/h to 2000 rem/h) |
| Hp(10) γ Display resolution dose rate | 1 μSv/h to 10.0 Sv/h (0.1 mrem/h to 1000 rem/h) up to three decimal places |
| Hp(0.07) γ Effective range of dose rate | 0.05 μSv/h to 20 Sv/h (0.005 mrem/h to 2000 rem/h) |
| Hp(0.07) γ Display resolution dose rate | 1 μSv/h to 10.0 Sv/h (0.1 mrem/h to 1000 rem/h) up to three decimal places |
| Photon Hp(10) On-axis energy response | ±15% from 15 keV to 1.5 MeV; -15% to +20% from 1.5 MeV to 10 MeV |
| Photon Hp(0.07) On-axis energy response | ±30% from 20 keV to 1.5 MeV; -15% to +20% from 1.5 MeV to 10 MeV |
| Photon Hp(10) Combined energy and angular response | -29% to +67% from 16 keV to 10 MeV, 0° to 60° |
| Photon Hp(0.07) Combined energy and angular response | -29% to 67% from 24 keV to 10 MeV, 0° to 60° |
| Photon Hp(10) Accuracy | ±5% |
| Photon Hp(0.07) Accuracy | ±5% |
| Power | |
| Battery | standard AAA (LR03) 1.5 V Alkaline |
| Physical | |
| Dimensions | 87 x 60 x 21 mm (3.4 x 2.3 x 0.8 in) max. without clip |
| Weight | < 88 g (3.1 oz) with alkaline battery and clip |
| Medical X-ray pulse width > 1 ms Max. pulse dose rate | 0.05 μSv/h to 5 Sv/h (0.005 mrem/h to 500 rem/h) |
| Medical X-ray pulse width > 1 ms Relative response | ±20% for pulse width >1 ms (-40% at 10 Sv/h, (1000 rem/h)) |
| Connectivity | |
| Data storage capacity | more than 2,500 steps |
| Other | |
| Autonomy | 12 calendar months battery life (typical, 8 h per day, 5 days per week in run mode, without excessive alarms); 3,000 h battery life in continuous mode, without excessive alarm |
| Case | rugged, high impact polycarbonate-ABS |
| Replaceable clips | 3 back clips and 1 front-facing clip |
| Relative humidity | < 90% at +42 °C (+108 °F) |
| Storage | -20 °C to +71 °C (-4 °F to +160 °F) without battery |
| Shock, vibration and drop resistant | 1.5 meter on concrete |
| Waterproof | IP67: 1 m (39 in) during 1 h |
| Dose increments steps | configurable steps (10 s, 60 s, 10 min, 1 h, 24 h) |
| Event log storage | alarms, faults, changes saved during the selected time periods |
| Communication frequency | 125 kHz |
| Hp(10) γ Overload indication | from 10 Sv to > 100 Sv (1000 rem to > 10000 rem) |
| Hp(0.07) γ Overload indication | from 10 Sv to > 100 Sv (1000 rem to > 10000 rem) |
| Hp(10) γ Overload indication dose rate | from 10 Sv/h to > 50 Sv/h (1000 rem/h to > 5000 rem/h) |
| Hp(0.07) γ Overload indication dose rate | from 10 Sv/h to > 50 Sv/h (1000 rem/h to > 5000 rem/h) |
| Dose rate linearity | ±10% Between 10 Sv/h (1000 rem/h) and 50 Sv/h (5000 rem/h) cumulative dose |
Calibration, repair and maintenance for your detection instruments.