The Mirion Technologies SABG-100 is a 100 cm² phoswich probe for direct measurement of alpha, beta, and gamma surface contamination in public safety surveys.
The SABG-100 probe for measurement of surface contamination is designed to be used with any CSP survey meter or any other device able to handle CSP protocol. Its phoswich detector with 100 cm² detection area makes it an ideal tool for direct measurement of alpha, beta and gamma emitters. The probe is part of the Canberra SMART Probe (CSP) family and includes all key components of hardware circuitry with intelligence located in the probe for controlling components, storing parameters, settings, calibrations, and alarm settings. Features include calibration via PC, easy removable grid for decontamination, ergonomic counting mode selector on probe body, and new probe angle for hands-free contamination measurement. The probe can store up to 1000 data points and supports multiple measurement units (CPM, DPM, DPM/100 cm², c/s, Bq, Bq/cm²). It connects via 1.5 meter or 20 meter CSP cable and can be upgraded via CSPS software.
| Performance | |
| Unit to display | Depending on survey meter (c/s, Bq, Bq/cm² or CPM, DPM, DPM/100 cm²) |
| Detector | ZnS(Ag) adhered to 1.5 mm thick plastic scintillation material |
| Detection area | 102 cm² (68 x 150 mm) |
| Measurement range | 0 to 10000 c/s, 0 to 600 kcpm. Activity equivalent range depends on calibration emitter. Conversion coefficients are factory set with ²³⁹Pu for alpha channel and with ⁶⁰Co for beta channel |
| Energy range | Beta > 150 keV, Alpha > 3 MeV, Gamma > 100 keV |
| Gamma sensitivity activity | Sealed source positioned at 20 mm from the protection grid |
| Display | Provided by survey meter |
| Alarm setpoints | 10 values for each unit to display. Saved in probe memory. They can be edited with CSPS an PC. Default alarm threshold is chosen in a list by use of survey meter keypad |
| Power | |
| Power | Supplied by survey meter or PC (low voltage only): +5 V |
| Physical | |
| Housing | painted aluminium |
| Dimensions | Length (with connector) x width (detector) x height (detector): 318.5 x 99 x 102 (12.5 x 3.9 x 4 in) |
| Weight | 710 g (25 oz) without cable |
| Other | |
| Emitters | Alpha and beta-gamma |
| Removable aluminized Mylar entrance window on metallic frame, thickness | 6 µm |
| Protection grid transparency | 83 % |
| Dead time | 2 µs |
| Gamma sentitivity in dose rate for ¹³⁷Cs | > 70 c/s per µGy/h (42 kcpm per mR/h) |
| Background | Ambient < :100 nSv/h (10 µR/h): Alpha <0.05 c/s (<3.0 cpm), beta < 10.0 c/s, (<600 cpm) |
| Sunlight effect | No effect up to 80 000 lux |
| Cross Talk | Alpha to Beta (²³⁹Pu) < 12 %, Beta to Alpha (⁹⁰Sr-⁹⁰Y) < 0.1 % |
| Consumption | 15 mA maximum |
| Temperature | -20 °C to +50 °C (-4 °F to 122 °F) use and storage |
| Relative humidity | 10% to 93% at 35°C |
| Cleaning | Housing easy to decontaminate |
| IP | IP20 |
| EMC | conforms |
| CE | meets CE requirements |
| IEC 60325 | meets standard requirements |
Calibration, repair and maintenance for your detection instruments.