The PM1703MA-II/BT is a personal radiation detector for first responders and security personnel that operates reliably in extreme temperatures with NORM suppression.
The PM1703MA-II/BT is a gamma-only Personal Radiation Detector (PRD) equipped with a high-sensitive CsI(Tl) SiPM scintillator for measurement of personal dose rate up to 300 µSv/h. Features ANSI N42.32-2016 and IEC 62401:2017 compliance, NORM-suppression algorithm, Bluetooth 4.0 communication with Polismart app, USB connectivity, audible/visual/vibration alarms, extreme temperature operation (-40°C to 50°C), IP65 hermetic case, and 1000-hour battery life. Designed for border control, HAZMAT teams, counter-terrorism, homeland security, and first responders.
| Performance | |
| Detector | CsI(Tl) SiPM |
| Dose rate measurement range | 0.1 µSv/h to 300 µSv/h |
| Dose rate measurement accuracy | ± 30% |
| Energy range | 33 keV to 3.0 MeV (search) |
| Gamma radiation sensitivity | 100 cps per µSv/h (137Cs) 500 cps per µSv/h (241Am) |
| Response time | 0.25 s |
| Alarms | visual, audible, vibration |
| Atmospheric pressure | 84 to 106.7 kPa |
| Power | |
| Power supply | one AA alkaline or rechargeable battery |
| Battery lifetime | 1000 h |
| Physical | |
| Dimensions | 87 × 72 × 32 mm |
| Weight | 200 g |
| Connectivity | |
| Memory | 2000 events |
| Approvals & Warranty | |
| Ingress protection | IP65 |
| Other | |
| Smartphone communication | Bluetooth 4.0 |
| PC communication | USB |
| Drop test | 1.5 m |
| Operating temperature | –40 °C to 50 °C |
| Humidity | up to 98 % at +35 °C |
| Mean service lifetime | 8 years |
Calibration, repair and maintenance for your detection instruments.