A bump test proves that gas reaches the sensors and the alarms go off. A calibration adjusts the sensors so the numbers on the screen are right. The safety-equipment industry's position is a check before each day's use, but full-calibration intervals come from each manufacturer, so this guide quotes what each one actually says.
People use "calibrate" for everything they do with a cylinder of gas. The International Safety Equipment Association (ISEA), the trade group for safety-equipment makers, splits the work into three tests in its Statement on Validation of Operation for Direct Reading Portable Gas Monitors. Knowing which one you are doing tells you what the result means.
A qualitative check. You flow gas over the sensors long enough, and at a high enough level, to trip at least the low alarm. It proves gas can reach the sensor and that every alarm works. Pass/fail is usually judged on response time or a minimum response, such as 80% of the gas applied. It does not tell you the meter is accurate.
A quantitative check with a known, traceable gas. The reading must land within the manufacturer's limits. ISEA says that is typically ±10–20% of the gas applied, unless the manufacturer, your company or a regulator sets a different window. Nothing is adjusted.
An adjustment. The instrument compares each sensor's response with a known, traceable gas and resets it to match. It is done the way the manufacturer's instructions describe. On instruments that track a calibration due date, this is the step that resets it.
Sensit says it plainly in the SENSIT GOLD 3 manual: a bump test is not a substitute for calibration, and it does not update the calibration due date. Only a full calibration resets the due date for each sensor.
A calibration sets two points on each sensor's curve. The zero point is what the sensor reads in clean air. The span point is what it reads on a known concentration of gas. Honeywell's MultiRAE and QRAE 3 manuals say to set zero before any other calibration. The MultiRAE, QRAE 3 and AreaRAE manuals all say to do it in clean air with 20.9% oxygen or with a cylinder of clean zero air. For the oxygen sensor, the fresh-air step sets the reading to the oxygen in ambient air, about 20.9%.
Zero is only as good as the air you zero in. If the "fresh" air holds a little of the gas you are measuring, that amount becomes the new zero and every later reading is low by the same amount. That is why Honeywell makes two exceptions on the MultiRAE family. A CO2 sensor must be zeroed with 100% nitrogen, because CO2 is always in ambient air. A parts-per-billion PID sensor should be zeroed through a charcoal filter, a VOC zeroing tube or a bottle of zero-grade air, because ambient air normally holds trace VOCs. Sensit gives the same rule for the GOLD 3: always start and zero it in a gas-free environment.
Every sensor type wears in its own way. Electrochemical cells slowly use up their chemistry. Catalytic-bead LEL sensors lose response when contaminants coat the bead. PID lamps lose output as the lamp window gets dirty, and Honeywell's PID technical note says readings vary with lamp age and cleanliness. Drift is gradual, so a meter can pass yesterday's bump and still read low today.
The ISEA statement lists the events that should trigger an extra check, beyond the daily one:
Honeywell's RAE manuals give the same triggers for a bump test: physical impact, liquid immersion, an Over Limit alarm, a change of custody, or any time the monitor's performance is in doubt. Sensit adds a warning for the LZ30 laser detector: calibration checks that keep coming in low can mean the sensor is near end of life or failing.
The industry-wide position comes from ISEA. Bump test or calibration-check every portable gas monitor before each day's use, following the manufacturer's instructions. Do a full calibration at regular intervals set by the manufacturer, your company or a regulator. ISEA does not set one interval for everyone, so the numbers below come from each manufacturer's own manual.
Bump and calibration guidance, quoted from each manufacturer's documentation
| Instrument | Bump test | Full calibration |
|---|---|---|
| Honeywell MultiRAE, MultiRAE Lite, MultiRAE Pro (one shared manual) | Before each day's use. Also after impact, liquid immersion, an Over Limit alarm, a custody change, or whenever performance is in doubt. | When it fails a bump test, when a new sensor is installed, after sensor maintenance, and at least once every 180 days, depending on use and exposure to poisons and contaminants. |
| Honeywell AreaRAE Plus and AreaRAE Pro | Before each day's use, with the same extra triggers. | Same rule: failed bump, new sensor, sensor maintenance, or at least once every 180 days. |
| Honeywell QRAE 3 | "Periodically" for the full bump test. The combustible sensor must be checked before each day's use on methane at 20–50% of full scale, reading within 0 to +20% of actual. | When it fails a bump test, when a new sensor is installed, or at least once every 180 days. |
| Honeywell Flex 5 Series | Scanning the BUMP tag on a gas cylinder starts a bump test with full traceability. The datasheet and brochure give no interval. | Scanning a CALIBRATION tag runs a guided calibration. The detector keeps the last 10 calibrations in its history. |
| Sensit GOLD 3 (G3) | Built-in bump test passes when each sensor reaches at least 80% of the gas value within a set time. A bump test does not reset the calibration due date. | At your company-specified interval. The due date and days remaining are shown at startup, and a "calibration past due" message appears when it is overdue. |
| Sensit LZ30 laser methane detector | Can be done before each use, using the gas cell built into the carry case. A pass is 1,500 ppm·m or more. | With the built-in cell only. The cell gas is certified for 3 years, then Sensit refills and re-certifies it. |
Source: Honeywell MultiRAE Series User Manual Rev L (M01-4003-000); AreaRAE Plus/Pro User Manual Rev G (W01-4001-000); QRAE 3 User's Guide Rev F (M02-4001-000); Honeywell Flex 5 Series datasheet (04/26); SENSIT GOLD 3 Instruction Manual 750-00094 (06/14/2022); GAS TRAC LZ30 Instruction Manual 750-00096 (12/09/2022)
Read the 180-day figure as a ceiling, not a target. The front pages of the MultiRAE and AreaRAE manuals say the same thing another way: calibrate every time the monitor fails a bump test, but no less often than every six months, depending on use, gas exposure and contamination. Both also say intervals can vary with national rules and company policy. A meter that lives on a hazmat rig and sees solvent vapor, silicone sprays or over-range hits will need calibration well before day 180.
Honeywell's RAE instruments can enforce your schedule. With Policy Enforcement on, a MultiRAE or QRAE 3 prompts when a bump or calibration is due. Set to "Can't Bypass", the MultiRAE cannot be used until the test is done. Set to "Can Bypass", it still runs, but it records the skipped test in a Policy Violation report.
ISEA's rule is short: a monitor that fails a bump test or calibration check must get a full calibration before further use, or come out of service. In practice, work through it in this order:
When good gas still will not calibrate a sensor, Honeywell's troubleshooting table points to a bad sensor or a bad pump. FarrWest's gas detector calibration and repair service uses OEM-certified technicians, with a typical 48–72 hour depot turnaround and calibration certificates tied to each instrument's serial number. For fleets, the on-site Service Program runs on a 30–180 day cycle.
A test you cannot prove is hard to defend after an incident. For each bump and calibration, log the date and time, the instrument serial number, who did it, each sensor's result and reading, the cylinder's lot number and expiry date, and what you did about any failure.
Docking stations make this easier. Honeywell's AutoRAE 2 handles charging, data management, bump testing and calibration for the MultiRAE family. Sensit's SMART-CAL 360 station syncs the GOLD 3's clock on every connection so the calibration and bump records across a fleet line up. The Flex 5 keeps its own history of the last 10 calibrations.
Match the gas to the sensors. Honeywell recommends cylinders that contain the gas each sensor needs, at the correct concentration. A MultiRAE with a PID plus toxic and LEL sensors usually needs two cylinders: isobutylene for the PID and a four-gas mix for the rest. The manual's four-gas example is 50% LEL methane, 18% O2, 10 ppm H2S and 50 ppm CO. Sensit's G3 bump test uses separate gases: 2.5% methane, 100 ppm CO and 25 ppm H2S in air, and 10 ppm HCN in nitrogen. The G3's oxygen sensor is checked with its own O2 test, not the bump test.
Watch the date. Every certified cylinder is labeled with its contents and an expiration date. Do not bump or calibrate with gas past that date. Sensit's LZ30 cell, for example, is certified for 3 years and is then considered expired.
Use the regulator the manual calls for. For a manual bump on a MultiRAE or QRAE 3, Honeywell specifies a constant-flow regulator at 0.5–1 liter per minute with the calibration adapter fitted. An AutoRAE 2 must use demand-flow regulators and no adapter. For the AreaRAE Plus/Pro, Honeywell ranks a 500 cc/min constant-flow regulator as good, 750 cc/min with a T-tube as better and demand flow as best. Honeywell also calls for Teflon tubing when testing or calibrating a PID sensor on the MultiRAE and AreaRAE.
Cylinders, regulators, zero air and nitrogen are in FarrWest's calibration equipment category.
A bump test is a quick functional check. It exposes the sensors to enough gas to set off the alarms and proves gas reaches the sensors. A calibration adjusts each sensor against a known, traceable gas so the readings are accurate. A bump test does not reset the calibration due date.
ISEA's position is a bump test or calibration check before each day's use, following the manufacturer's instructions. Honeywell's MultiRAE and AreaRAE manuals say the same, and also call for a bump after impact, liquid immersion, an Over Limit alarm, a custody change or whenever performance is in doubt.
Honeywell says the MultiRAE must be calibrated when it fails a bump test, when a new sensor is installed, after sensor maintenance, and at least once every 180 days, depending on use and exposure to poisons and contaminants. Heavy exposure means calibrating sooner.
Take it out of use. Check the calibration gas, regulator and adapter first, let the sensors warm up and retest. If it still fails, calibrate the failed sensors. If calibration fails, remove the instrument from service and send it for repair.
Yes for most sensors, as long as the air is truly clean, with 20.9% oxygen. Honeywell's exceptions are CO2 sensors, which need 100% nitrogen, and parts-per-billion PID sensors, which need a charcoal filter, VOC zeroing tube or zero-grade air.
No. A bump test only proves the sensors respond and the alarms work. Accuracy needs a calibration check against a known gas, and adjustment needs a full calibration.
OEM-certified technicians, 48–72 hour typical depot turnaround, and certificates tied to each serial number, or on-site service every 30 to 180 days.