A photoionization detector (PID) calibrated on isobutylene reads every vapor as if it were isobutylene. A correction factor (CF) converts that number into the compound you are actually measuring: multiply the reading by the CF. This guide shows the math, explains lamp energy and blind spots, and lists Honeywell's published CFs for common hazmat and industrial compounds.
A PID pulls air past an ultraviolet lamp. The lamp's photons knock an electron off any molecule that needs less energy to ionize than the photons carry. The charged fragments are collected on electrodes, and the current they make rises with concentration. The instrument turns that current into ppm or ppb.
That design makes a PID fast and very sensitive to volatile organic compounds (VOCs). It also makes it non-specific. The PID cannot tell benzene from acetone. It reports the combined signal from everything in the sample that its lamp can ionize. Honeywell's Technical Note TN-106 puts the rule simply: in general, any compound with an ionization energy below the lamp's photon energy can be measured.
Different compounds give different signals at the same concentration. Ten ppm of toluene makes a much bigger signal than 10 ppm of ethanol. The most accurate way to calibrate is with the exact gas you expect to measure, but nobody carries a cylinder of every VOC.
So PIDs are calibrated on one reference gas, typically isobutylene, and the reading is in isobutylene units. Honeywell has measured how hundreds of other compounds respond relative to isobutylene; the MultiRAE ships with a copy of TN-106 covering 300+ VOCs. Those ratios are the correction factors, and they let one cylinder of isobutylene stand in for a long list of chemicals. Isobutylene's own CF is 1.00 by definition.
The rule is one line: actual concentration = reading × CF, with the instrument calibrated on isobutylene. A CF below 1 means the PID responds more strongly to that compound than to isobutylene, so the raw reading runs high. A CF above 1 means a weaker response, so the raw reading runs low. A very large CF, in the hundreds, means the lamp barely sees the compound. Any small error in the reading is then multiplied by that large number.
Example 1, toluene. A MultiRAE calibrated on isobutylene reads 20 ppm in a toluene spill area. Honeywell's MultiRAE CF for toluene on a 10.6 eV lamp is 0.53. 20 × 0.53 = about 11 ppm toluene.
Example 2, n-hexane. The same meter reads 20 ppm near a hexane solvent tank. The MultiRAE CF is 3.4. 20 × 3.4 = 68 ppm hexane. The raw reading was less than a third of the real number.
TN-106 gives three ways to apply a CF on Honeywell PIDs:
Converting to mg/m³. TN-106 gives a short formula for air at 25 °C (77 °F): mg/m³ = ppm × molecular weight × 0.041. For 50 ppm toluene (molecular weight 92.1), that is 50 × 92.1 × 0.041 = about 189 mg/m³.
Every compound has an ionization energy (IE), measured in electron-volts (eV). Older documents call it ionization potential; TN-106 notes that "ionization energy" is the more correct term. Every PID lamp has a photon energy, also in eV. If the lamp's energy is above the compound's IE, the PID can generally see it. Pick the lamp, and you pick what the PID can and cannot see.
Sees aromatics such as benzene (IE 9.25) and toluene (8.82) well, but responds weakly or not at all to many higher-IE compounds. In TN-106, n-hexane (IE 10.13) has a CF of 350 on a 9.8 eV lamp. Honeywell offers a 9.8 eV PID on the MultiRAE and MultiRAE Pro; with it, those models can take snapshot benzene readings using RAE-Sep tubes in Benzene Mode.
The standard lamp on the MiniRAE 3000+ and in the MultiRAE family's ppm and ppb PID sensors. In TN-106 it has listed CFs for fuels, aromatics, ketones, alcohols and chlorinated solvents, as the table below shows.
Ionizes compounds that a 10.6 eV lamp misses. In TN-106, methanol (IE 10.85) shows no response on 9.8 and 10.6 eV lamps but has a CF of 2.5 on 11.7 eV. Chlorine (IE 11.48) has a CF of 1.0 on 11.7 eV and no listed value on the lower lamps. The MiniRAE 3000+ datasheet lists an optional 11.7 eV lamp.
A lamp change is a calibration event. Honeywell's MultiRAE manual flags a sensor as needing calibration when the lamp type has been changed.
The IE is too high. Methane (IE 12.61), hydrogen cyanide (13.6) and carbon monoxide (14.01) are above even an 11.7 eV lamp, so a PID gives no response. The main components of clean air are not ionized either, which is why a PID reads near zero in fresh air. This is why a PID never replaces an LEL sensor for methane or electrochemical sensors for CO and HCN.
The response is too weak. Some compounds are ionized but give a small signal. TN-106 lists ammonia at a 10.6 eV CF of 10.9, and Honeywell's MultiRAE-specific note lists it at 32.26. On a MultiRAE, 10 ppm on the display could mean about 320 ppm ammonia. A dedicated ammonia sensor, such as the MultiRAE NH3 sensor, is the better tool. TN-106 also notes that some inorganic compounds, such as hydrogen peroxide and nitrogen dioxide, respond weakly even when their IE is well below the lamp energy.
The vapor never reaches the lamp. TN-106 warns that high-boiling ("heavy") compounds may not vaporize enough to register, even when their IE is low enough. Heavy or very reactive compounds can also stick to filters, pumps and other sensors in the sample path. Honeywell's MultiRAE note says multi-gas meters such as the MultiRAE and AreaRAE Plus/Pro place the pump and other sensors upstream of the PID. For those compounds, the reading may be slow, or missing in extreme cases, and may only give a rough relative indication.
Correction factors are lab numbers. TN-106 says they were measured at room temperature in what it calls dry air (40–50% relative humidity), typically at 50–100 ppm. They apply from the ppb range up to about 1,000 ppm. Above that, the CF may change, and Honeywell recommends calibrating with the actual gas near the concentration you expect.
Mixtures. If you know the vapor composition, TN-106 gives a formula for a mixture CF: CFmix = 1 ÷ (X1/CF1 + X2/CF2 + …), where X is each compound's fraction of the vapor. Take a vapor that is 5% benzene and 95% n-hexane on a MultiRAE (CFs 0.48 and 3.4). CFmix = 1 ÷ (0.05/0.48 + 0.95/3.4) = about 2.6. A reading of 100 means about 260 ppm total vapor: about 13 ppm benzene and 248 ppm hexane. If you do not know what is in the air, there is no valid CF to apply. Treat the reading as an isobutylene-equivalent screening number and identify the vapor with other tools.
Honeywell's MultiRAE, MultiRAE Lite, MultiRAE Pro, AreaRAE Plus and AreaRAE Pro do the math for you. In the menu, "Change Measurement Gas" offers four lists:
Once you pick a gas, the monitor applies its CF and reads in that compound. On a MultiRAE with five sensors including a PID, the screen shows the lamp, the CF in use and the measurement gas. A Toggle Display option, turned on in SSDC, alternates between the LEL and VOC readings and their CFs. Honeywell says the CFs in Technical Note TN-106B match the MultiRAE Series library in firmware V1.52.
Three details trip people up. First, during a manual bump test or calibration, the MultiRAE shows readings in units of the calibration gas, not the measurement gas. Second, changing the calibration gas type without recalibrating flags the sensor as needing calibration. Third, for better accuracy at high concentrations, the MultiRAE and MultiRAE Pro can run a three-point PID calibration on 10.6 eV sensors. It is off by default and is enabled in SSDC; the MultiRAE Lite's PID does not support it.
FarrWest's photoionization detectors category lists the PID instruments we carry, including Honeywell's ModRAE modular platform.
The table below pulls 25 compounds from two Honeywell technical notes. TN-106 is Honeywell RAE Systems' general guideline for its PIDs. TN-106B lists the CFs built into the MultiRAE Series library. Honeywell says the factors are tuned to each instrument, so the same compound can have different CFs on different models. TN-106B also says its CFs do not apply to other manufacturers' instruments.
PID correction factors (multiply an isobutylene-calibrated reading by the CF) and ionization energies
| Compound | IE (eV) | CF 9.8 eV (TN-106) | CF 10.6 eV (TN-106) | CF 11.7 eV (TN-106) | CF 10.6 eV, MultiRAE library (TN-106B) |
|---|---|---|---|---|---|
| Isobutylene (reference gas) | 9.24 | 1.00 | 1.00 | 1.00 | 1 |
| Benzene | 9.25 | 0.55 | 0.47 | 0.6 | 0.48 |
| Toluene | 8.82 | 0.54 | 0.45 | 0.51 | 0.53 |
| Ethylbenzene | 8.77 | 0.52 | 0.47 | 0.51 | 0.77 |
| m-Xylene | 8.56 | 0.50 | 0.44 | 0.40 | 0.44 |
| Styrene | 8.43 | 0.45 | 0.43 | 0.4 | 0.6 |
| Acetone | 9.71 | 1.2 | 0.9 | 1.4 | 1.08 |
| Methyl ethyl ketone (MEK) | 9.51 | 0.86 | 0.8 | 1.1 | 1.1 |
| Ethanol | 10.47 | — | 7.9 | 3.1 | 11 |
| Isopropanol (IPA) | 10.12 | 500 | 4.2 | 2.7 | 3.3 |
| n-Hexane | 10.13 | 350 | 5.1 | 0.51 | 3.4 |
| 1,3-Butadiene | 9.07 | 0.8 | 0.6 | 1.1 | 0.73 |
| Vinyl chloride | 9.99 | — | 2.0 | 0.6 | 2 |
| Trichloroethylene (TCE) | 9.47 | 0.62 | 0.54 | 0.43 | 0.54 |
| Perchloroethylene (PCE) | 9.32 | 0.69 | 0.57 | 0.31 | 0.57 |
| Gasoline (#2, 92 octane) | — | 1.3 | 1.0 | 0.5 | 1 |
| Diesel fuel #2 | — | 1.3 | 0.7 | 0.4 | 0.7 |
| Jet fuel JP-8 | — | — | 0.94 | 0.3 | 0.6 |
| Ethylene oxide | 10.57 | — | 13 | 3.5 | 13 |
| Ammonia | 10.16 | NR | 10.9 | 5.7 | 32.26 |
| Hydrogen sulfide | 10.45 | NR | 3.3 | 1.5 | 3.3 |
| Methanol | 10.85 | NR | NR | 2.5 | ND |
| Methane | 12.61 | NR | NR | NR | ND |
| Hydrogen cyanide | 13.6 | NR | NR | NR | not listed |
| Carbon monoxide | 14.01 (TN-106B) | not listed | not listed | not listed | ND |
Source: Honeywell RAE Systems Technical Note TN-106, "A Guideline for PID Instrument Response" (11/18/VK), ; Honeywell Technical Note TN-106B, "Correction Factors, Ionization Energies, and Calibration Characteristics for MultiRAE Series Monitors" (202005/TM), . NR = no response (TN-106). ND = not detectable (TN-106B). — = no value listed. Gasoline row = TN-106 "Gasoline #2, 92 octane" and TN-106B "Gasoline". Diesel row = TN-106 "Diesel Fuel #2 (Automotive)" and TN-106B "Diesel Fuel #2, whole". CO's IE is from TN-106B; TN-106 does not list CO.
Look at ammonia, ethanol and ethylbenzene. The general TN-106 values and the MultiRAE library values differ a lot. Use the CF your instrument applies, which is the one shown on its screen, rather than a number from another chart. When accuracy matters for one specific compound, TN-106 recommends calibrating with that compound.
It is the number you multiply an isobutylene-calibrated PID reading by to get the concentration of the compound you are actually measuring. It reflects how strongly the PID responds to that compound compared with isobutylene.
Calibrate on isobutylene, then multiply the reading by the compound's CF. For example, 20 ppm on a MultiRAE in toluene vapor × 0.53 = about 11 ppm toluene. Or select the compound in the monitor's gas library, and it applies the CF for you.
Calibrating with every possible compound is impractical. Honeywell has measured correction factors relative to isobutylene for hundreds of compounds, so one isobutylene cylinder lets a PID quantify many chemicals.
No. Methane (IE 12.61 eV) and carbon monoxide (14.01 eV) have ionization energies above the 9.8, 10.6 and 11.7 eV lamps. Use an LEL sensor for methane and an electrochemical sensor for CO.
The lamp energy sets which compounds the PID can ionize. A 9.8 eV lamp is more selective, a 10.6 eV lamp is the general-purpose standard, and an 11.7 eV lamp reaches compounds such as methanol and chlorine that a 10.6 eV lamp cannot.
Honeywell says no for its MultiRAE Series: its CFs are tuned to its own instruments and do not apply to other manufacturers' PIDs. Even between Honeywell's own notes, values differ, so use the CF your instrument applies.
OEM-certified technicians, 48–72 hour typical depot turnaround, and certificates tied to each serial number, or on-site service every 30 to 180 days.